Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85
- Recall number
- Z-1503-2014
- Initiated
- March 17, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips Medical Systems, Inc.
- Quantity
- 985 devices
App-derived interpretation
The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
Code information
Units distributed from November 2006 through March 2014.
Distribution pattern
Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico