Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67787

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 19, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista¿ System. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.

Z-1647-2014
Recall number
Z-1647-2014
Initiated
March 19, 2014
Classification
Class II
Status
Terminated
Quantity
4175

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C

Code information

CSA 13114BA, exp. 2014-04-24; CSA 13157BA, exp. 2014-06-06; CSA 13246BB, exp. 2014-09-03; CSA 13317BA, exp. 2014-11-13; CSA 14008BB, exp. 2015-01-08; CSA 14050BB, exp. 2015-02-19; CSAE 13157BC, exp. 2014-06-06; CSAE 13249BD, exp. 2014-09-06.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Belgium, Switzerland, Germany, Spain, France, Great Britain, Italy, Saudi Arabia, Australia, and Japan.