Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67719

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 24, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Brainlab AG

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED

Z-1513-2014
Recall number
Z-1513-2014
Initiated
February 24, 2014
Classification
Class II
Status
Terminated
Recalling firm
Brainlab AG
Quantity
41 systems (13 systems in US; 28 systems to foreign countries)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.

Code information

ExacTrac version v.6.0.3.

Distribution pattern

Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France, Germany, Japan, Netherlands, and the United Kingdom.