Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67513

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 01, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Abbott Point Of Care Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.

Z-1306-2014
Recall number
Z-1306-2014
Initiated
January 01, 2014
Classification
Class II
Status
Terminated
Recalling firm
Abbott Point Of Care Inc.
Quantity
48, 450 US, 1400 OUS (49,850 catridges)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.

Code information

List number 03P78-25 :Lot number N13183

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada.

Field note

Send feedback