Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67342

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 14, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Instrumentarium Dental, PaloDEx Group Oy

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The FOCUS is a microprocessor controlled intra-oral x-ray unit with a Load Matching Frequency (LMF) DC generator, which produces dental images on digital and film media. For the digital images the FOCUS is connected to the Sigma USB terminal which is connected to a PC, which reads the image data from a sensor and displays the image on the PC monitor. In order to make installation and service faster and easier the electronics is integrated into the horizontal arm. FOCUS is intended to be used for producing diagnostic x-ray radiographs of dentition, jaws and other oral structures.

Z-1091-2014
Recall number
Z-1091-2014
Initiated
January 14, 2014
Classification
Class II
Status
Terminated
Quantity
101 units (US Distribution)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Instrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were delivered with FOCUS X-ray machines that had not been properly welded at our supplier. As a result of the incomplete welding operation, the wall mount unit on the FOCUS X-ray unit may not properly support the unit as expected, which could cause the unit to drift. Depending on the extent of the shortened or missing welds, there is also a possibility that the wall mount unit will fail to support the weight of the unit, resulting in the FOCUS X-ray unit falling unexpectedly from its mounted position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Instrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were delivered with FOCUS X-ray machines that had not been properly welded at our supplier. As a result of the incomplete welding operation, the wall mount unit on the FOCUS X-ray unit may not properly support the unit as expected, which could cause the unit to drift. Depending on the extent of the shortened or missing welds, there is also a possibility that the wall mount unit will fail to support the weight of the unit, resulting in the FOCUS X-ray unit falling unexpectedly from its mounted position.

Code information

Device Model Number / Sales Code: 50600-IMG; Serial Numbers (US and Foreign): F21221 - F21641, and F21643; Serial Numbers (US Devices): F21221 - F21230, F21265 - F21270, F21273 - F21290, F21296 - F21350, F23154 - F21365, F21367 - F21391, F21395 - F21447, F21451 - F21464, F21469 - F21478, F21481 - F21486, F21490 - F21519, F21525 - F21542, F21545 - F21550, F21565 - F21588, F21623 - F21628, F21636 - F21641, and F21643

Distribution pattern

Worldwide Distribution: US (Nationwide) including states of: AK, AZ, CA, CO, GA, IL, KY, MD, MT, NC, NV, OR, PA, TN, TX, UT, VA, WA,and WI; and Internationally to: Belgium Brazil, Chile, China, Czech Republic, France, Germany, Italy, Spain, Sweden, Switzerland, and United Kingdom.

Field note

Send feedback

We'll only use this to respond to your feedback.