Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67319

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 13, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Healthcare

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908

Z-0845-2014
Recall number
Z-0845-2014
Initiated
January 13, 2014
Classification
Class II
Status
Terminated
Recalling firm
Philips Healthcare
Quantity
50

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.

Code information

Model Numbers: 861470 (120V), 861472 (240V).

Distribution pattern

Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.

Field note

Send feedback

We'll only use this to respond to your feedback.