Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66609

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 24, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ossur North America Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Fixed Offset Adapter, A-642100. The Offset Adapter is a fixed modular component which sets the tube 14.5 mm (0.57 inch) offset from the center line of the pyramid. It is made of a high-strength alloy with a pyramid receiver and is intended where there is a need to offset the tube horizontally in relation to other components. It is fully compatible with all standard 30mm prosthetic pylons. (3) Offset: 14,5mm (0.57inch).

Z-0362-2014
Recall number
Z-0362-2014
Initiated
October 24, 2013
Classification
Class II
Status
Terminated
Recalling firm
Ossur North America Inc
Quantity
66 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ossur inititated a voluntary recall of the A-642100 Fixed Offset Adapter because it can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ossur inititated a voluntary recall of the A-642100 Fixed Offset Adapter because it can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely.

Code information

Item Number: A-642100 All Serial Numbers.

Distribution pattern

US Distribution including the states of AL, TX, NY, CA, VA, GA, WI, NC, ID and NJ.

Field note

Send feedback

We'll only use this to respond to your feedback.