Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66352

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 02, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Instruments Div. of Stryker Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Z-0092-2014
Recall number
Z-0092-2014
Initiated
October 02, 2013
Classification
Class II
Status
Terminated
Quantity
67

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

Code information

10933, 11433, 11483, 20553, 30163, 31553, 40713, 40833, 41113, 50463, 60133, 60423, 90383, 110093, 110193, 1030133, 1030403, 1040413, 2010013, 2010123, 96090403, 96091343, 97020193, 97030013, 97030543, 97050243, 97060163, 97080033, 98010083, 98020703, 98030103, 98040213, 98060023, 98061033, 98070603, 98090023, 98090473, 98090503, 98101143, 98110323, 99010433, 99030253, 99050033, 99050383, 99060163, 99060423, 99061063, 99061283, 99070453, 99071293, 99080243, 99080483, 99090913, 99101273, 221303733, 326102033, 434909453, 500800103, 504900233, 631112793, 631414853, 700314923, 723606673, 725315933, 735502183, 813304543, and 921700523 ,

Distribution pattern

Worldwide distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Italy, Japan, Spain, Switzerland, and United Kingdom.

Field note

Send feedback