Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66204

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 10, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Biomet 3i, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Osseotite 2 Certain Prevail 5mm(D)X 4.1mm (P) X X 11.5mm (L) Dental Implant Product Usage: Surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.

Z-2207-2013
Recall number
Z-2207-2013
Initiated
June 10, 2013
Classification
Class II
Status
Terminated
Recalling firm
Biomet 3i, LLC
Quantity
215

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

Code information

XIIOS5411

Distribution pattern

Worldwide Distribution - USA Nationwide in the states of MA, CT, NJ, DE, NY, FL, OK, TX, PA, DC, IL, KS, LA, MD, CA, RI, TN, MN, and SC and the countries of Canada, Japan, Australia, Italy Germany, France, Switzerland, Apain, Brazil, Uruguay, Singapore, Dominican Republic, Taiwan, Austria, Portugal, and the United Kingdom.

device · product 2 of 2

Osseotite Parallel Walled Certain 4mm(D)X 15mm (L) Dental Implant Product Usage: Surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.

Z-2208-2013
Recall number
Z-2208-2013
Initiated
June 10, 2013
Classification
Class II
Status
Terminated
Recalling firm
Biomet 3i, LLC
Quantity
178

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

Code information

IOSS415

Distribution pattern

Worldwide Distribution - USA Nationwide in the states of MA, CT, NJ, DE, NY, FL, OK, TX, PA, DC, IL, KS, LA, MD, CA, RI, TN, MN, and SC and the countries of Canada, Japan, Australia, Italy Germany, France, Switzerland, Apain, Brazil, Uruguay, Singapore, Dominican Republic, Taiwan, Austria, Portugal, and the United Kingdom.

Field note

Send feedback

We'll only use this to respond to your feedback.