Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
- Recall number
- Z-2303-2013
- Initiated
- October 08, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Radiometer America Inc
- Quantity
- 13,632 boxes/4 membranes
App-derived interpretation
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Code information
Part Number: 942-066 R1741 to R1869
Distribution pattern
Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.