Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66101

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 26, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Bayer Healthcare Llc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.

Z-0136-2014
Recall number
Z-0136-2014
Initiated
August 26, 2013
Classification
Class II
Status
Terminated
Recalling firm
Bayer Healthcare Llc
Quantity
138 Kits in US (Total 2452)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.

Code information

Lot: 1300733 (US, China); Lot: 1300712 (Argentina, Uruguay); Lots: 1300711, 130012 (Japan)

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to China, Argentina, Uruguay, Japan.

Field note

Send feedback

We'll only use this to respond to your feedback.