Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65644

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 03, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic MiniMed

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic MiniMed Paradigm Reservoirs, Model No: MMT-326A (1.8mL) and MMT-332A (3.0mL) The model MMT-326A and MMT-332A are syringe type insulin reservoirs intended for use with Medtronic Paradigm series insulin infusion pumps. The two reservoirs are identical except for barrel length and as a result, maximum insulin capacity. The MMT-326A holds a maximum of 1.8ml of insulin whereas the longer MMT-332A reservoir can hold up to 3.0ml of insulin. Both reservoirs are supplied sterile and are labeled for single use. Typically, the reservoir is replaced every tow or three days at the time of infusion set replacement. The Paradigm reservoirs have a shelf life of three years post sterilization. Product Quantity Distributed (Int) 669,672 units

Z-1986-2013
Recall number
Z-1986-2013
Initiated
July 03, 2013
Classification
Class II
Status
Terminated
Recalling firm
Medtronic MiniMed
Quantity
669,672 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic is recalling certain lots of Medtronic MiniMed Paradigm Reservoirs MMT-326A (1.8mL) and MMT-332A (3.0mL) used with Medtronic Paradigm insulin pumps because they may have increased risk for leaking. A leak in the reservoir may result in delivery of less insulin than intended. In addition, if there is a leaky reservoir and an insulin blockage occurs in the infusion set, the pump may not

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic is recalling certain lots of Medtronic MiniMed Paradigm Reservoirs MMT-326A (1.8mL) and MMT-332A (3.0mL) used with Medtronic Paradigm insulin pumps because they may have increased risk for leaking. A leak in the reservoir may result in delivery of less insulin than intended. In addition, if there is a leaky reservoir and an insulin blockage occurs in the infusion set, the pump may not

Code information

MMT-326A Lot Numbers: H8437486 H8441420 H8442973 H8451531 H8486688 H8489386 H8485398 H8500423 H8500472 H8503728 H8512826 H8503372 H8510440 H8539013 H8627745 MMT-332A Lot Numbers: H8416432 H8420977 H8422490 H8424676 H8452933 H8455959 H8457716 H8459557 H8461538 H8463297 H8464121 H8467888 H8469703 H8471745 H8473271 H8476270 H8478398 H8473106 H8492449 H8491921 H8494645 H8496561 H8509305 H8512566 H8515317 H8517079 H8521052 H8541843 H8584244 H8603292 H8604958 H8635301

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Australia, Bermuda, Brazil, Canada, Chile, Columbia, Ecuador, El Salvador, Germany, Japan, Mexico, Netherland, Paraguay, Puerto Rico, Singapore, Uruguay and Venezuela.

Field note

Send feedback

We'll only use this to respond to your feedback.