Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65194

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 08, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Intuitive Surgical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.

Z-1431-2013
Recall number
Z-1431-2013
Initiated
May 08, 2013
Classification
Class II
Status
Terminated
Recalling firm
Intuitive Surgical, Inc.
Quantity
697 units - all products

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

Code information

Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M10121115 M10121123 M10121130 M10121206 M10121210 M10130124 M10130131 M10130205 M10130304 M10130319 M10130321

Distribution pattern

Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.

device · product 2 of 3

8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.

Z-1432-2013
Recall number
Z-1432-2013
Initiated
May 08, 2013
Classification
Class II
Status
Terminated
Recalling firm
Intuitive Surgical, Inc.
Quantity
697 - all products

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

Code information

Model number: 400172-08; Lot numbers: M10120815 M10120817 M10120831 M10120913 M10120918 M10121008 M10121024 M10121120 M10121123 M10121203 M10121210 M10121215 M10130220 M10130222 M10130311 M10130315 M10130327 M10130411.

Distribution pattern

Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.

device · product 3 of 3

8mm Fenestrated Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.

Z-1433-2013
Recall number
Z-1433-2013
Initiated
May 08, 2013
Classification
Class II
Status
Terminated
Recalling firm
Intuitive Surgical, Inc.
Quantity
697 - all products

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

Code information

Model number: 400205-04; Lot numbers: M10120816 M10120820 M10120903 M10120910 M10120912 M10120921 M10121008 M10121024 M10121120 M10121130 M10121206 M10121213 M10121215 M10130124 M10130131 M10130213 M10130218 M10130222 M10130313 M10130319 M10130402.

Distribution pattern

Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.