device · product 1 of 3
8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
- Recall number
- Z-1431-2013
- Initiated
- May 08, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 697 units - all products
App-derived interpretation
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Code information
Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M10121115 M10121123 M10121130 M10121206 M10121210 M10130124 M10130131 M10130205 M10130304 M10130319 M10130321
Distribution pattern
Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.