Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64578

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 08, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Spinal Solutions, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

APLIF Implants and Instruments, Part Numbers: 184008-10 Anterior Psoas Lumbar 18 x 40 x 08mm - 10degree 184010-10 Anterior Psoas Lumbar 18 x 40 x 10mm - 10degree 184012-10 Anterior Psoas Lumbar 18 x 40 x 12mm - 10degree 184014-10 Anterior Psoas Lumbar 18 x 40 x 14mm - 10degree 184016-10 Anterior Psoas Lumbar 18 x 40 x 16mm - 10degree 184008-00 Anterior Psoas Lumbar 18 x 40 x 08mm - 00degree 184010-00 Anterior Psoas Lumbar 18 x 40 x 10mm - 00degree 184012-00 Anterior Psoas Lumbar 18 x 40 x 12mm - 00degree 184014-00 Anterior Psoas Lumbar 18 x 40 x 14mm - 00degree 184016-00 Anterior Psoas Lumbar 18 x 40 x 16mm - 00degree 184510-08 Anterior Psoas Lumbar 18 x 45 x 08mm - 10degree 184510-10 Anterior Psoas Lumbar 18 x 45 x 10mm - 10degree 184512-10 Anterior Psoas Lumbar 18 x 45 x 12mm - 10degree 184514-10 Anterior Psoas Lumbar 18 x 45 x 14mm - 10degree 184516-10 Anterior Psoas Lumbar 18 x 45 x 16mm - 10degree 184510-00 Anterior Psoas Lumbar 18 x 45 x 10mm - 00degree 184512-00 Anterior Psoas Lumbar 18 x 45 x 12mm - 00degree 184514-00 Anterior Psoas Lumbar 18 x 45 x 14mm - 00degree 184516-00 Anterior Psoas Lumbar 18 x 45 x 16mm - 00degree 185008-10 Anterior Psoas Lumbar 18 x 50 x 08mm - 10degree 185010-10 Anterior Psoas Lumbar 18 x 50 x 10mm - 10degree 185012-10 Anterior Psoas Lumbar 18 x 50 x 12mm - 10degree 185014-10 Anterior Psoas Lumbar 18 x 50 x 14mm - 10degree 185016-10 Anterior Psoas Lumbar 18 x 50 x 16mm - 10degree 185008-00 Anterior Psoas Lumbar 18 x 50 x 08mm - 00degree 185010-00 Anterior Psoas Lumbar 18 x 50 x 10mm - 00degree 185012-00 Anterior Psoas Lumbar 18 x 50 x 12mm - 00degree 185014-00 Anterior Psoas Lumbar 18 x 50 x 14mm - 00degree 185016-00 Anterior Psoas Lumbar 18 x 50 x 16mm - 00degree 185508-10 Anterior Psoas Lumbar 18 x 55 x 08mm - 10degree 185510-10 Anterior Psoas Lumbar 18 x 55 x 10mm - 10degree 185512-10 Anterior Psoas Lumbar 18 x 55 x 12mm - 10degree 185514-10 Anterior Psoas Lumbar 18 x 55 x 14mm - 10degree 185516-10 Anterior Psoas Lumbar 18 x 55 x 16mm - 10degree 185508-00 Anterior Psoas Lumbar 18 x 55 x 08mm - 00degree 185510-00 Anterior Psoas Lumbar 18 x 55 x 10mm - 00degree 185512-00 Anterior Psoas Lumbar 18 x 55 x 12mm - 00degree 185514-00 Anterior Psoas Lumbar 18 x 55 x 14mm - 00degree 185516-00 Anterior Psoas Lumbar 18 x 55 x 16mm - 00degree The Eminent Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment.

Z-1157-2013
Recall number
Z-1157-2013
Initiated
March 08, 2013
Classification
Class II
Status
Terminated
Recalling firm
Spinal Solutions, LLC
Quantity
220 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
inadequate sterilization

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

No Marketing Application

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could cause patient harm due to implant breakage, movement, or inadequate sterilization.

Code information

All lots are subject to this recall.

Distribution pattern

Nationwide Distribution including NV, WI, MD,and CA.

Field note

Send feedback

We'll only use this to respond to your feedback.