Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64220

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 19, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Hospira Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.

Z-1016-2017
Recall number
Z-1016-2017
Initiated
February 19, 2013
Classification
Class II
Status
Terminated
Recalling firm
Hospira Inc.
Quantity
48,050 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.

Code information

Product List Number: 14699-28; Lot Number: 17-077-NS

Distribution pattern

Nationwide Distribution

Field note

Send feedback

We'll only use this to respond to your feedback.