device · product 1 of 1
Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1 in human plasma. Catalog # 101201. Spectrolyse PAI-1 is intended for the quantitative determination of Plasminogen Activator Inhibitor Type-1 (PAI-1) activity in human plasma. The test is for in vitro diagnostic use.
Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.
These labels are deterministic app interpretations, not FDA categories.
Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.
Code information
Lot # 120827, exp 11/27/2012
Distribution pattern
Nationwide Distribution-including the states of FL, IN, NC, OH, UT, and WA.