Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63577

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 01, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Roche Diagnostics Operations, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.

Z-0473-2013
Recall number
Z-0473-2013
Initiated
November 01, 2012
Classification
Class II
Status
Terminated
Quantity
78 analyzers

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C

Code information

Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff¿ Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08.

Distribution pattern

Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.

Field note

Send feedback

We'll only use this to respond to your feedback.