Timeline bucket
August 09, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Spine

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Stryker XIA 3 TORQUE WRENCH REF 48237028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Z-0380-2013
Recall number
Z-0380-2013
Initiated
August 09, 2012
Classification
Class II
Status
Terminated
Recalling firm
Stryker Spine
Quantity
54

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Code information

Class I device, 510(K) exempt Catalog Number: 48237028 Lot Number: 093653, 093654, 093655

Distribution pattern

Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

device · product 2 of 3

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Z-0381-2013
Recall number
Z-0381-2013
Initiated
August 09, 2012
Classification
Class II
Status
Terminated
Recalling firm
Stryker Spine
Quantity
19 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Code information

Class I device, 510(K) exempt Catalog Number: 48287028 Lot Number: 098400

Distribution pattern

Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

device · product 3 of 3

SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Z-0382-2013
Recall number
Z-0382-2013
Initiated
August 09, 2012
Classification
Class II
Status
Terminated
Recalling firm
Stryker Spine
Quantity
28 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Code information

Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756

Distribution pattern

Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

Field note

Send feedback