Event 63383

5 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 19, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Sorin Group USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

5 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 5

S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0271-2013
Recall number
Z-0271-2013
Initiated
September 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
35 units

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi

Code information

Serial Numbers: 10E06219,10E06220, 10E06231-10E06234, 10E06239-10E06241, 10E06255-10E06273, 10E06286-10E06289, 10E06291-10E06355, 10E06359-10E06364, 10E06366, 10E06368-10E06383, 10E06385, 10E06387-10E06391,10E06395-10E06401, 10E06403-10E06447, 10E6502.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.

device · product 2 of 5

S5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0272-2013
Recall number
Z-0272-2013
Initiated
September 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
5 units

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Code information

Serial Numbers: 10E51854,10E51856-10E51858, 10E51860, 10E51861, 10E51864, 10E51867, 10E51870, 10E51871, 10E51873, 10E51880-10E51897, 10E51899-10E51913.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.

device · product 3 of 5

S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0273-2013
Recall number
Z-0273-2013
Initiated
September 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
3 units

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Code information

Serial Numbers: 50E50766--50E50769, 50E50772-50E50774, 50E50783-50E50784.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.

device · product 4 of 5

S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0274-2013
Recall number
Z-0274-2013
Initiated
September 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
2 units

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Code information

Serial Numbers: 50E00680--50E00681, 50E00688, 50E00696-50E00706, 50E00708-50E00718.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.

device · product 5 of 5

Sorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0275-2013
Recall number
Z-0275-2013
Initiated
September 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
2 units (no product distributed within the US).

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Code information

Serial Numbers: 58E00135; 58E00136

Distribution pattern

Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.