Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63353

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 14, 2007
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Advanced Bionics Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument.

Z-0117-2013
Recall number
Z-0117-2013
Initiated
September 14, 2007
Classification
Class II
Status
Terminated
Quantity
17 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.

Code information

Serial Numbers: 351323, 251153, 350989, 351226, 350196, 351136, 350140, 350201, 351261, 351264, 351266, 351269, 351412, 351415, 351429, 350157, 351303.

Distribution pattern

Worldwide distribution: USA (nationwide) including state: IL and country of: Europe.