Event 62790

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 03, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories

Z-2245-2012
Recall number
Z-2245-2012
Initiated
August 03, 2012
Classification
Class II
Status
Terminated
Quantity
126 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.

Code information

System Code # 728243, Serial #s: 7392, 7394, 7399, 7405, 7406, 7429, 7433, 7440, 7457, 7465, 7469, 7479, 7481, 7487, 7488, 7491, 7502, 7506, 7507, 7517, 7518, 7520, 7521, 7532, 7539, 7545, 7547, and 7556. System Code # 728244, Serial #s: 7190, 7195, 7318, 7383, 7427, 7430, 7510, and 7533. System Code #728256, Serial #s: 3163, 3500, 3500, 5623, 5655, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5813, 5836, 5855, 5862, 5868, 5877, 5897, 5911, 5911, 5916, 5935, 5969, and 5997. System Code #728251, Serial #: 4051. System Code # 728246, Serial #s: 5547, 5585, 5791, 5803, 5876, 5974. System Code # 728235, Serial #s: 9233, 9234, 9261, 95022. System Code # 728231, Serial #s: 9867, 9925, 9931, 9980, 9996, 10009, 10011, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10133, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10153, 10166, 10170, 10172, 10198, 10203, 95109, 95348, 95563.

Distribution pattern

Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, DC, FL, GA, KY, MD, MI, MN, MS, ND, NM, NY, OH, OK, SC, TN, TX, VT, WA, and WI and the countries of: Austria, China, Denmark, France, Germany, Iceland, Japan, Netherlands, South Africa, Sweden, Switzerland, United Kingdom. There were two government consignees. There were no Canadian consignees.