Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62535

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 19, 2012
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Impact Instrumentation, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Model 731 EMV+ Uni-Vent, Impact Instrumentation Inc., 27 Fairfield Place, West Caldwell NJ 07006. Indicated for use in the management of acute or chronic respiratory failure or during resuscitation by providing continuous positive -pressure ventilation.

Z-2152-2012
Recall number
Z-2152-2012
Initiated
March 19, 2012
Classification
Class III
Status
Terminated
Quantity
Ten Units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ten units of Model EMV+ Devices were shipped out in error before the verification and validation was completed for a new integrated circuit (IC) on the CPU PCB.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ten units of Model EMV+ Devices were shipped out in error before the verification and validation was completed for a new integrated circuit (IC) on the CPU PCB.

Code information

Serial numbers: VCDB120315, VCDB120316, VCDB120299, VCDB120300, VCDB120313, VCDB120307, VCDB120312, VCDB120303, VCDB120310, VCDB120297.

Distribution pattern

US Distribution to the state of Illinois.

Field note

Send feedback

We'll only use this to respond to your feedback.