Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62481

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 08, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips And Neusoft Medical Systems Co., Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".

Z-2083-2012
Recall number
Z-2083-2012
Initiated
June 08, 2012
Classification
Class II
Status
Terminated
Quantity
15 units distributed in the US

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.

Code information

Serial No.: N16E0 90002, N16E0 90003, N16E0 90004, N16E0 90007, N16E0 90008, N16E0 90010, N16E1 00015, N16E1 00017, N16E1 00023, N16E1 10016, N16E1 10023, N16E1 10026, N16E1 10027, N16E1 10040, and N16E1 10053.

Distribution pattern

Nationwide Distribution -- North Carolina, Ohio, and Texas.

Field note

Send feedback

We'll only use this to respond to your feedback.