Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62137

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 05, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.

Z-1933-2012
Recall number
Z-1933-2012
Initiated
June 05, 2012
Classification
Class II
Status
Terminated
Recalling firm
Zimmer, Inc.
Quantity
77 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we

Code information

Part number 3564-1; all lots

Distribution pattern

Worldwide distribution: USA (nationwide) and countries of: Australia, France and Germany.

Field note

Send feedback

We'll only use this to respond to your feedback.