Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 37530

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 06, 2006
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Radiometer America Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Z-0608-2013
Recall number
Z-0608-2013
Initiated
June 06, 2006
Classification
Class III
Status
Terminated
Recalling firm
Radiometer America Inc
Quantity
840 Ampules

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Code information

Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Distribution pattern

Distributed to KY, NC, and NY.

Field note

Send feedback

We'll only use this to respond to your feedback.